Joanne Hancox, David Spalton, Catherine Heatley, Hari Jayaram, Jennifer Yip, James Boyce, John Marshall
Journal: Journal of cataract and refractive surgery 2007;33(3):413-7
PMID: 17321391
PURPOSE
To measure posterior capsule opacification (PCO) and neodymium:YAG (Nd:YAG) capsulotomy rates between the AcrySof MA30 intraocular lens (IOL) (Alcon) and the 1CU IOL (HumanOptics) in a fellow-eye comparison.
SETTING
Ophthalmology Department, St. Thomas' Hospital, London, United Kingdom.
METHODS
Thirty patients who had bilateral cataract surgery with a 1CU IOL prospectively randomly allocated to 1 eye and an AcrySof MA30 monofocal IOL to the other eye were examined. Best corrected distance visual acuity was recorded using the Early Treatment Diabetic Retinopathy Study logMAR chart. Digital retroillumination images of the posterior capsule were taken with the pupil dilated and analyzed with POCO software.
RESULTS
Eyes with the 1CU IOL had significantly higher PCO rates than eyes with the MA30 IOL at all time points. By 2 years after surgery, 50% of eyes with a 1CU IOL had required Nd:YAG capsulotomy compared with no eyes with an MA30 IOL. There was no significant difference in visual acuity at any time point when post Nd:YAG capsulotomy was taken in to account.
CONCLUSIONS
The 1CU IOL has 4 broad optic-haptic junctions where the square-edged barrier is breached; this appeared to allow passage of lens epithelial cells, leading to an increase in PCO. However, the increased PCO cannot be attributed to this alone as the 1CU is hydrophilic, a factor known to be associated with higher PCO rates.