Temperature as a variable in liquid chromatography: development and application of a model for the separation of model drugs using water as the eluent.

A M Edge, S Shillingford, C Smith, R Payne, I D Wilson

Journal: Journal of chromatography. A 2006;1132(1-2):206-10

PMID: 16934820

Abstract

The use of high temperatures in liquid chromatography allows for the use of a purely aqueous mobile phase. At elevated temperatures water possesses many of the characteristics of organic solvents in terms of eluotropic strength, as well as having a lower viscosity. A model is developed, based on data obtained using a range of model drugs, which demonstrates the relationship between temperature, flow and pressure. Experimental data from different column types, at temperatures from 40 degrees C to 180 degrees C, is presented which matches well with the predicted data from the model.

Address: Department of Drug Metabolism and Pharmacokinetics, AstraZeneca Pharmaceuticals, Alderley Park, Macclesfield, Cheshire SK10 4TG, UK. [email protected] <[email protected]>