Solid state "self-calibrated" EPR-dosimeters-advantageous and shortcomings.

Z Peimel-Stuglik, S Fabisiak

Journal: Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy 2006;63(4):855-60

PMID: 16495136

Abstract

EPR dosimetry method with alpha-alanine as radiation sensitive material (RSM) is widely used in high dose dosimetry laboratories. However, it is not suitable for routine industrial applications mainly because of difficult EPR measurement procedure. In order to simplify quantitative EPR dosimetry measurements Yordanov and Gancheva developed so-called "self-calibrated" (sc) dosimeters consists of RSM (alpha-alanine, sugar, other ones), Mn2+/MgO as internal EPR intensity standard (IES) and a binder. The aim of this work was to check dosimetric properties of two experimental batches of sc EPR dosimeters with alpha-alanine and sugar as RSM, Mn2+/MgO as IES and paraffin as a binder. The percentage content of the components was 60, 5 and 35% (w/w), respectively. It was established that the investigated alanine sc-dosimeters are about two times more sensitive than the sugar ones. The dose-response coefficient, K(dr) of sc-alanine dosimeters was stable in all investigated dose range from 1 to 23 kGy. The K(dr)-value of sugar sc-dosimeters decreased with the dose what was in a contradiction to the results pointed to the high stability of radiation generated sugar radicals. The observed effect arose probably from the special chemical procedure used for the sc-sugar dosimeters production. The results confirmed an expectation that the position of sc dosimeter in the cavity is not important factor for accurate dose evaluation. It allows to read-out dosimetric signals in shorter time, with lower uncertainty and on less sophisticate EPR-spectrometers than that commonly used till now. The main shortcomings of sc dosimeters are: (a) the limitation of RSM suitable for sc dosimeters to these ones having strictly linear signal to dose characteristic; (b) necessity to assure very good homogeneity of dosimeter material; and (c) the cost of IES present in the amount of some percent in each sc dosimeter.

Address: Laboratory for Measurement of Technological Doses, Institute of Nuclear Chemistry and Technology, Dorodna 16, 03-195 Warsaw, Poland. [email protected]

Link outs

Subscription / membership required

Bant logo

© Copyright 2026, Nutrition Evidence

NED wishes to thank the following organisations for their support:

We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.