Comparison of cardiac troponin I levels in serum and sodium-citrate plasma using the ACCESS 2 immunoassay.

Li-Yu Tsai, Yu-Sin Wang, Chen-Fen Feng, Tze-Kiong Er, Chian-Yu Chen, Shun-Jen Chang, Bai-Hsiun Chen

Journal: Annals of clinical and laboratory science 2006;35(4):453-4

PMID: 16254264

Abstract

The aim of this study was to determine if sodium-citrate plasma is a suitable substitute for serum for cardiac troponin I (cTnI) measurement using the Beckman ACCESS 2 immunoassay. Linear regression analysis of cTnI levels in paired serum and sodium-citrate plasma samples from 80 patients provided a y-intercept of -0.0193 and a slope of 1.044. The correlation was good [r=0.995; 95% confidence interval 1.017 to 1.068, p <0.0001]. The paired results were not significantly different by Student's t-test. Serum and sodium-citrate plasma appear to be interchangeable as samples for cTnI measurement using the ACCESS 2 assay. Furthermore, the use of plasma reduces the turnaround time and may avoid false positive cTnI results due to residual fibrin.

Address: Department of Laboratory Medicine, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan.

Link outs

Free resources

Subscription / membership required

Bant logo

© Copyright 2026, Nutrition Evidence

NED wishes to thank the following organisations for their support:

We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.