Carlos De Sadeleer
Journal: Nuclear medicine review. Central & Eastern Europe 2005;8(1):15-20
PMID: 15977142
BACKGROUND
Conflicting results have been published concerning the reproducibility of Tc-99m MAG3 clearance. The aim of the study was to reevaluate again this reproducibility on a large prospective series of healthy volunteers.
MATERIAL AND METHODS
Fifty subjects underwent three successive tests performed at 1-week interval. The physiological conditions were controlled as much as possible and all experimental measurements were rigorously double-checked. Renal clearances were calculated using single sample algorithms.
RESULTS
Thin layer chromatography demonstrated a radiochemical purity of more than 92% (mean 93.8%; SD 1.6%; range 92.0-96.4%). The mean (and SD) of 44th minute plasma concentrations for the three successive measurements were respectively 1.8 +/- 0.4, 2.1 +/- 0.4 and 1.9 +/- 0.4 (%dose/liter). The mean changes (and S.D. of differences) between two tests were 0.26 +/- 0.29, -0.08 +/- 0.5 and 0.18 +/- 0.30 respectively between tests 1 and 2, tests 1 and 3, and tests 2 and 3. Using Bubeck's algorithm, the mean clearance values (ml/min/1.73 m2) were 268.8 (range 201.4-336.8), 247.2 (range 170.5-290.3), and 262.8 (range 187.4-340.0) respectively for the first, second and third measurements. Using Russell's approach, the mean clearance values were respectively 314.9 (range: 208.7-441.7), 280.1 (range: 167.5-358.6), and 305.4 (range: 189.5-447.9).
CONCLUSIONS
High differences were observed between the 3 tests. Using Russell s formula, the SD of differences between two tests was respectively 35.8, 47.7 and 57.7 ml/min/1.73 m2 between tests 1 and 2, tests 1 and 3, and tests 2 and 3. Whether such a large variability is acceptable in clinical practice depends solely on what the clinician is expecting.
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.