BACKGROUND
Intracoronary radiation therapy (IRT) prevents recurrent in-stent restenosis, but its long-term safety and efficacy remain uncertain. In the present study, the long-term clinical outcome of IRT using the rhenium-188 ((188)Re)-filled balloon system was evaluated.
METHODS AND RESULTS
After successful catheter-based treatment of either a de novo or restenotic lesion, 187 patients were randomly assigned to either the radiation (N=104) or the control (N=83) group. The (188)Re-filled balloon system was designed to deliver 17.6 Gy to 1.0-mm tissue depth. Angiographic restenosis was significantly reduced with IRT at 9 months (18.9% vs 45.9%, p<0.001), but the incidence of major adverse cardiac events (MACE) including death, myocardial infarction, and target-vessel revascularization (TVR) by 3 years showed no difference. Lack of clinical benefit might be related to TVR caused by geographic miss (6/22, 28.6%), balloon-induced unhealed dissection (3/22, 13.6%) and late thrombosis (2/22, 9.1%). In the restenotic subgroup (N=39), the MACE rate within 3 years was significantly reduced with IRT (14.3% vs 54.5%, p=0.01).
CONCLUSIONS
IRT using the (188)Re -filled balloon system is safe and technically feasible. Although IRT failed to show favorable outcomes for de novo lesion, the clinical benefits for restenotic lesions seem durable for 3 years. Furthermore, preventing geographic miss and dissection might improve long-term outcomes.
Full Text Sources:
Medical:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.