Kuan-Hsun Huang, Lung-Ming Fu, Chin-Chung Tseng, Chia-Chun Lee, Cheng-Xue Yu
Journal: Biosensors 2025;15(2):
PMID: 39997006
A microfluidic system for detecting sodium ions (Na) has been developed, incorporating a micro finger-pump chip and a micro-spectrometer platform to measure Na concentration in human serum. A small volume (10 μL) of serum sample is introduced into the microchip and reacted with a preloaded reagent mixture through a two-step finger-pump actuation process. The resulting purple complex is directed into the detection area of the chip and analyzed using the micro-spectrometer at wavelengths of 555 and 666 nm. The Na concentration is then inversely derived from the measured A/A absorbance ratio using self-written software installed on a Raspberry Pi. The entire detection process is completed in just 3 min, offering a significant advantage in meeting clinical needs compared to the traditional reporting turnaround time of several hours in medical institutions. The experimental results indicate a linear relationship between the measured absorbance ratio and Na concentration within the range of 1-200 mM, with a correlation coefficient of R = 0.9989. Additionally, the detection results from 60 serum samples collected from chronic kidney disease (CKD) patients showed a strong agreement with those obtained using the conventional indirect ion-selective electrode (ISE) method, achieving a correlation coefficient of R = 0.9885 and an average recovery rate of 99.4%. In summary, the proposed system provides a practical, affordable, and rapid alternative to conventional Na detection methods, making it highly promising for point-of-care (POC) testing applications.
Full Text Sources:
Miscellaneous:
Research Materials:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.