Woo Jin Bang, Jae Hoon Chung, Deok Hyun Han
Journal: Investigative and clinical urology 2025;66(2):137-143
PMID: 40047127
PURPOSE
To compare encrustation resistance between silicon- and polytetrafluoroethylene (PTFE)-covered metallic ureteral stents (MUS) in an infection model and to determine the most effective material for reducing biofilm formation and encrustation.
MATERIALS AND METHODS
A total of 52 MUS were prepared: 26 silicon-covered and 26 PTFE-covered stents. Each sample was immersed in artificial urine inoculated with in a biofilm reactor for 48 hours. After immersion, the stents were weighed to measure their encrustation level. Scanning electron microscopy (SEM) and energy dispersive X-ray spectroscopy (EDS) were used to assess the surface morphology and elemental composition of the encrustation deposits.
RESULTS
Silicon-covered stents showed a statistically significant reduction in weight gain due to encrustation compared to PTFE-covered stents (9.50±5.77 mg vs. 16.75±10.61 mg; p=0.004). Additionally, encrustation per unit length was lower in silicon-covered stents (0.76±0.45 mg/mm vs. 1.30±0.81 mg/mm; p=0.004). SEM and EDS analyses demonstrated lower calcium salt deposition on the silicon-covered stents, indicating greater resistance to encrustation.
CONCLUSIONS
Silicon-covered MUS demonstrated superior resistance to encrustation compared to PTFE-covered stents, supporting silicon as a more suitable covering material for long-term MUS applications. This finding may lead to extended stent lifespans and a reduced frequency of stent replacements, benefiting both patients and healthcare systems.
© The Korean Urological Association.
Full Text Sources:
Molecular Biology Databases:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.