Barbara Dziedzic, Janina Mazanowska-Gajdowicz, Anna Walczewska, Agata Sarniak, Dariusz Nowak
Journal: Clinica chimica acta; international journal of clinical chemistry 2004;335(1-2):65-74
PMID: 12927686
BACKGROUND
Analysis of NO2-/NO3- in expired breath condensate (EBC) has been proposed as a marker of inflammation in various lung diseases.
METHODS
NO2- and total NO3-/NO2- concentrations were determined in EBC collected from healthy and asthmatic subjects. The NO3- was first reduced to NO2-, and total NO2- was detected by colorimetric Griess reaction. Two methods of NO3- reduction were compared. To reduce NO3-, cadmium (600 microl EBC-macromethod) and enzyme-NADPH-nitrate reductase (60 microl EBC-micromethod) were used.
RESULTS
Macromethod: Mean NO2- concentrations in EBC were 1.64 +/- 0.24 micromol/l in healthy subjects and 0.42 +/- 0.17 micromol/l in asthmatic patients. Mean total NO2-/NO3- levels were 3.64 +/- 0.43 micromol/l in healthy subjects and 3.27 +/- 0.34 micromol/l in asthmatic. Micromethod: NO2- level: 1.69 +/- 0.23 micromol/l in healthy subjects and 0.53 +/- 0.21 micromol/l in asthmatics. Total NO2-/NO3- levels: 3.56 +/- 0.37 micromol/l in healthy subjects and 3.57 +/- 1.17 micromol/l in asthmatics. Variability index was 27% and 6% for macro- and micromethod, respectively. Recovery of NO3- added to EBC was 100% for enzymatic and almost 88% for cadmium reduction. There was no correlation between total NO2-/NO3- levels determined by macro- and micromethod.
CONCLUSIONS
We recommend enzymatic reduction as a better method for NO3- determination in EBC.
Full Text Sources:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.