Geoffrey S Payne, Martin O Leach, Laura Mancini
Journal: Magnetic resonance in medicine 2004;50(3):578-88
PMID: 12939766
Several (31)P MRS studies in tumors in vivo have shown that levels of phosphocholine (PC) and other phosphomonoesters (PME) and phosphodiesters (PDE) are useful prognostic or early-response indicators. To improve sensitivity for such measurements, four polarization transfer (PT) sequences (insensitive nuclei enhanced by PT (INEPT), distortionless enhancement by PT (DEPT), reverse-INEPT, and heteronuclear single-quantum coherence (HSQC)) were assessed theoretically and experimentally. The presence of homonuclear ((1)H-(1)H) and heteronuclear ((31)P-(1)H) couplings of similar magnitude makes theoretical analysis very sensitive to precise model parameters, especially for the (1)H-detected sequences. The (1)H-(1)H coupling causes the splitting of (1)H peaks, and hence reduces the proton spectral resolution. This effect and a 50% signal loss from gradient-enhanced water suppression negate the usual advantages of (1)H-detection. Among the PT methods, INEPT gave the higher signal enhancement. However, T(2) losses during the long echo times (TEs) required by the weak coupling limited the resulting signals from PC.
Copyright 2003 Wiley-Liss, Inc.