Accuracy of rapid lateral flow immunoassays for human leptospirosis diagnosis: A systematic review and meta-analysis.

Supawadee Naorungroj, Teerapat Nualnoi, Luelak Lomlim

Journal: PLoS neglected tropical diseases 2024;18(5):e0012174

PMID: 38748731

Abstract

BACKGROUND

In the last two decades, several rapid lateral flow immunoassays (LFIs) for the diagnosis of human leptospirosis were developed and commercialized. However, the accuracy and reliability of these LFIs are not well understood. In this study, we aimed to evaluate the accuracy of leptospirosis LFIs as well as the factors affecting the test efficiency using systematic review and meta-analysis.

METHODS AND RESULTS

Original articles reporting the accuracy of human leptospirosis LFIs against microagglutination tests (MAT) or immunofluorescent assays (IFA) were searched from PubMed, Embase, and Scopus, and selected as per pre-set inclusion and exclusion criteria. A total of 49 data entries extracted from 24 eligible records published between 2003 and 2023 were included for meta-analysis. A meta-analysis was performed using STATA. The quality of the included studies was assessed according to the revised QUADAS-2. Only nine studies (32.1%) were considered to have a low risk of bias and no concern for applicability. Pooled sensitivity and specificity were calculated to be 68% (95% confidence interval, CI: 57-78) and 93% (95% CI: 90-95), respectively. However, the ranges of sensitivity (3.6 - 100%) and specificity (53.5 - 100%) of individual entries are dramatically broad, possibly due to the heterogeneity found in both study designs and LFIs themselves. Subgroup analysis demonstrated that IgM detection has better sensitivity than detection of IgG alone. Moreover, the test performance seems to be unaffected by samples from different phases of infection.

CONCLUSIONS

The pooled specificity of LFIs observed is somewhat acceptable, but the pooled sensitivity is low. These results, however, must be interpreted with caution because of substantial heterogeneity. Further evaluations of the LFIs with well-standardized design and reference test will be needed for a greater understanding of the test performance. Additionally, IgM detection type should be employed when leptospirosis LFIs are developed in the future.

Copyright: © 2024 Nualnoi et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Address: Department of Pharmaceutical Technology, Faculty of Pharmaceutical Sciences, Prince of Songkla University, Hat Yai, Songkhla, Thailand.; Drug Delivery System Excellence Center (DDSEC), Faculty of Pharmaceutical Sciences, Prince of Songkla University, Hat Yai, Songkhla, Thailand.; Department of Pharmaceutical Chemistry, Faculty of Pharmaceutical Sciences, Prince of Songkla University, Hat Yai, Songkhla, Thailand.; Phytomedicine and Pharmaceutical Biotechnology Excellence Center (PPBEC), Faculty of Pharmaceutical Sciences, Prince of Songkla University, Hat Yai, Songkhla, Thailand.; Department of Conservative Dentistry, Faculty of Dentistry, Prince of Songkla University, Hat Yai, Songkhla, Thailand.