Evaluation of Ultrathin Strut Biodegradable Polymer-Coated Sirolimus-Eluting Stents in an All-Comers Patient Population: 1-Year Results of the S-FLEX Slovakia Registry.

Martin Hudec, Andrej Kupec, Pavol Gazdic

Journal: Anatolian journal of cardiology 2024;28(3):142-149

PMID: 38419511

Abstract

BACKGROUND

Supraflex (Sahajanand Medical Technologies Limited, Surat, India) is a new-generation, biodegradable polymer-coated sirolimus-eluting stent (SES) designed on an ultrathin (60 µm) cobalt-chromium platform with a flexible 'S-link.' The S-FLEX Slovakia registry aimed to assess the safety and effectiveness of Supraflex SES in an all-comers population, with a subgroup of diabetic patients.

METHODS

This was a prospective, observational, multi-center, post-market registry conducted between February 2018 and May 2019. All consecutive patients with symptomatic coronary artery disease scheduled for percutaneous coronary intervention with Supraflex SES were enrolled. The primary endpoint was target lesion failure (TLF), defined as a composite of cardiac death, target vessel myocardial infarction (TV-MI), or clinically indicated target lesion revascularization (CI-TLR) by percutaneous or surgical methods at 1-year follow-up. Stent thrombosis was a safety endpoint.

RESULTS

A total of 413 patients was assessed (145 diabetics and 268 nondiabetics). At 1-year follow-up, the primary endpoint of TLF occurred in 5.1% patients, comprised of 3.9% cardiac deaths, 0.5% TV-MI, and 0.7% CI-TLR. Overall stent thrombosis occurred in 0.5% patients at 1-year follow-up. In the subgroup analysis, TLF occurred in 6.2% diabetics and 4.5% nondiabetics (P =.433) and comprised 4.8% and 3.4% cardiac deaths (P =.447), 0.7% and 0.4% TV-MI (P =.653), and 0.7%, and 0.7% CI-TLR (P =.952) in diabetics and non-diabetics, respectively. Overall stent thrombosis occurred in 0.7% diabetic and 0.4% non-diabetic patient (P =.659).

CONCLUSION

This registry demonstrates favourable clinical outcomes after the implantation of the ultrathin biodegradable polymer coated Supraflex SES in an all-comers population, with event rates that were similar in diabetic and nondiabetic patients.