Replacement gastrostomy tube causing acute pancreatitis: case series with review of literature.

Anish M Shah, Nihar Shah, Joseph R DePasquale

Journal: JOP : Journal of the pancreas 2012;13(1):54-7

PMID: 22233947

Abstract

CONTEXT

Percutaneous endoscopic gastrostomy (PEG) feedings are generally considered safe with few serious complications. Acute pancreatitis is a rare complication associated with replacement percutaneous endoscopic gastrostomy tubes.

CASE REPORT

We report two cases of acute pancreatitis induced by migrated replacement percutaneous endoscopic gastrostomy tubes.

CONCLUSIONS

Migration of a balloon into the duodenum can result in external manipulation of the ampulla of Vater thereby disturbing the flow of pancreatic secretions leading to acute pancreatitis. Recognition of this complication is important and should be included as potential etiology of acute pancreatitis in patients receiving percutaneous endoscopic gastrostomy feedings. Periodic examination and documentation of the distance of the balloon from the skin should be performed to document the position of the tubes or any inadvertent migration of the tubes. The use of Foley catheters as permanent replacement tubes should be considered medically inappropriate.

Address: Division of Gastroenterology, School of Health and Medical Sciences, Seton Hall University, South Orange, NJ 07079, USA. [email protected]