Dimitri Guy Rohner, Lukas Burget, Christoph Henzen, Stefan Fischli
Journal: Swiss medical weekly 2024;154():3366
PMID: 38579290
To avoid and reduce diabetic complications and glucose fluctuations (hypo- and hyperglycaemia) people with diabetes mellitus type 1 depend on close monitoring of their glucose values. The Eversense® CGM System is the first and only continuous glucose monitoring system (CGMS) that uses a fully subcutaneous implanted sensor. The aim of this study was to evaluate the effectiveness, safety, and patient-reported outcomes in patients with type 1 diabetes mellitus using the Eversense®CGM System. This study was a prospective and retrospective study, conducted at a single Swiss diabetes centre (Luzerner Kantonsspital). It included patients who had at least one Eversense® glucose sensor implanted between 2017 and 2022. Results showed that: - HbA1c levels changed by –0.25% at 6 months, –0.45% at 12 months, and –0.2% at the last follow-up after sensor implantation. - Premature sensor breakdowns occurred in 9% of cases, mainly between 2019 and 2020. - Major complications were rare, except for one infection and four complicated removals. - Patient-reported outcomes showed positive impacts on hypoglycaemia rates, confidence in managing hypoglycaemia, and diabetes management. Authors concluded that The Eversense® CGM System demonstrated favourable effects on HbA1c levels and patient-reported outcomes, with low rates of complications. Technical issues were reported but were rare with the newest sensor generation.
AIMS OF THE STUDY
The Eversense® CGM System is the first and only continuous glucose monitoring system (CGMS) that uses a fully subcutaneous implanted sensor. This study aimed to evaluate effectiveness, safety and patient-reported outcomes in patients using the Eversense® CGM System in a realistic clinical setting, assessed at a single Swiss diabetes centre (Luzerner Kantonsspital) with prolonged follow-up.
METHODS
This was a prospective and retrospective observational study that included patients with type 1 diabetes mellitus in whom at least one Eversense® glucose sensor was implanted between 2017 and 2022. The primary endpoint was the change in HbA1c levels from the baseline (before implantation of the sensor) to 6 ± 2 and 12 ± 2 months and the last follow-up (newest available value) after implantation. The secondary outcome measures were the number of premature sensor breakdowns, adverse events related to the implantation procedure (infection, bleeding, difficulties with implantation or explantation) and patient-related outcomes (assessed with a questionnaire).
RESULTS
A total of 33 patients participated in this study. The median follow-up time was 50 (IQR 22.3-58.5) months. In total, 178 sensor implantations were performed. Valid HbA1c results were available for 26 participants. Compared to the baseline values, HbA1c levels at 6 and 12 months and the last follow-up changed by -0.25%, -0.45 and -0.2 (p = 0.278, 0.308 and 0.296, respectively). We recorded 16 (9%) premature sensor breakdowns, all occurring between 2019 and 2020. Apart from one late-onset infection and four complicated sensor removals, no major complications were assessed. The results of the questionnaire showed a subjective improvement in hypoglycaemia rates, a better perception of hypoglycaemia and the impression of better diabetes management. Common issues with the device reported by the patients were technical errors (connection problems) and problems with the removal procedure.
CONCLUSIONS
The use of the Eversense® CGM System resulted in changes in HbA1c of between -0.2% and -0.45%. The rate of premature sensor breakdown was low. Major complications following sensor implantation or removal were absent, apart from one case of infection and four cases of complicated removal. Patient-reported outcomes with the Eversense® CGM System showed a subjective positive impact on hypoglycaemia rates, greater confidence in managing hypoglycaemia and diabetes in general, and easy handling of the transmitter and mobile app. Technical issues must be considered but are nowadays, with the use of the newest sensor generation, very rare.