Precision of slot widths and torque transmission of in-office 3D printed brackets : An in vitro study.

Carolien A J Bauer, Mats Scheurer, Christoph Bourauel, J Philippe Kretzer, Christoph J Roser, Christopher J Lux, Lutz D Hodecker

Journal: Journal of orofacial orthopedics = Fortschritte der Kieferorthopadie : Organ/official journal Deutsche Gesellschaft fur Kieferorthopadie 2024;85(5):327-339

PMID: 36862188

Abstract

PURPOSE

To investigate a novel in-office three-dimensionally (3D) printed polymer bracket regarding slot precision and torque transmission.

METHODS

Based on a 0.022″ bracket system, stereolithography was used to manufacture brackets (N = 30) from a high-performance polymer that met Medical Device Regulation (MDR) IIa requirements. Conventional metal and ceramic brackets were used for comparison. Slot precision was determined using calibrated plug gages. Torque transmission was measured after artificial aging. Palatal and vestibular crown torques were measured from 0 to 20° using titanium-molybdenum (T) and stainless steel (S) wires (0.019″ × 0.025″) in a biomechanical experimental setup. The Kruskal-Wallis test with post hoc test (Dunn-Bonferroni) was used for statistical analyses (significance level p < 0.05).

RESULTS

The slot sizes of all three bracket groups were within the tolerance range according to DIN 13996 (ceramic [C]: 0.581 ± 0.003 mm; metal [M]: 0.6 ± 0.005 mm; polymer [P]: 0.581 ± 0.010 mm). The maximum torque values of all bracket-arch combinations were above the clinically relevant range of 5-20 Nmm (PS: 30 ± 8.6 Nmm; PT: 27.8 ± 14.2 Nmm; CS: 24 ± 5.6 Nmm; CT: 19.9 ± 3.8 Nmm; MS: 21.4 ± 6.7 Nmm; MT: 16.7 ± 4.6 Nmm).

CONCLUSIONS

The novel, in-office manufactured polymer bracket showed comparable results to established bracket materials regarding slot precision and torque transmission. Given its high individualization possibilities as well as enabling an entire in-house supply chain, the novel polymer brackets bear high potential of future usage for orthodontic appliances.

© 2023. The Author(s).

Address: Poliklinik für Kieferorthopädie, Universitätsklinikum Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. [email protected].; Klinik und Poliklinik für Mund‑, Kiefer‑, Gesichtschirurgie, Universitätsklinikum Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.; Oralmedizinische Technologie, Zentrum für Zahn‑, Mund- und Kieferheilkunde, Universitätsklinikum Bonn, Welschnonnenstr. 17, 53111, Bonn, Germany.; Labor für Biomechanik und Implantatforschung, Klinik für Orthopädie, Universitätsklinikum Heidelberg, Schlierbacher Landstraße 200a, 69118, Heidelberg, Germany.; Poliklinik für Kieferorthopädie, Universitätsklinikum Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.