Adam B Yanke, Sachin Allahabadi, Zachary Wang, Kevin T Credille, Elizabeth Shewman, Joao Artur Bonadiman, Tristan J Elias, Mario Hevesi, Grant E Garrigues, Nikhil N Verma
Journal: The American journal of sports medicine 2023;51(10):2642-2649
PMID: 37345254
BACKGROUND
All-suture anchors and knotless anchors are increasingly used in the repair of anteroinferior labral tears in patients with shoulder instability. Optimal repair constructs may limit recurrent instability.
PURPOSE
To perform a quantitative biomechanical comparison of 3 labral fixation devices for soft tissue Bankart lesions: knotless soft-body tensionable anchor (SB knotless), knotted soft-body anchor (SB knotted), and knotless hard-body PEEK (polyether ether ketone) interference anchor (HB knotless).
STUDY DESIGN
Controlled laboratory study.
METHODS
A total of 21 glenoid specimens were randomized into 3 groups: SB knotless, SB knotted, and HB knotless. Artificial Bankart lesions were created at the anteroinferior labrum. Anchors were placed at the 3:30, 4:30, and 5:30 clockface positions, and sutures were passed through 1 cm of tissue. Anchors were tested simultaneously as one construct by pulling capsular tissue connected to the anteroinferior quadrant. Cyclic loading (5-25 N, 100 cycles) was followed by load-to-failure testing (15 mm/min). Biomechanical testing variables were collected, and failure mechanisms were recorded per individual anchor.
RESULTS
There were no differences in baseline specimen characteristics. There was no difference in elongation during cyclic loading ( = .40). The ultimate load to failure between SB knotless (309.7 ± 125.6 N), SB knotted (226.4 ± 34.8 N), and HB knotless (256.5 ± 90.5 N) did not significantly differ ( = .25). Failure mechanisms differed among groups ( = .008); mechanisms included anchor pullout (SB knotless: 33.3%; SB knotted: 23.8%; HB knotless: 28.6%), suture pull-through (SB knotless: 66.7%; SB knotted: 38.1%; HB knotless: 33.3%), and anchor fixation method failure, defined as knot failure for knotted anchors or locking mechanism failure for knotless anchors (SB knotless: 0.0%; SB knotted: 38.1%; HB knotless: 38.1%).).
CONCLUSION
The SB knotless, SB knotted, and HB knotless labral fixation anchors studied exhibited comparable elongation during cyclic loading, stiffness, and ultimate loads to failure in a cadaveric model. However, the failure mechanisms significantly differed, as SB knotless anchors failed primarily from suture pull-through, while SB knotted and HB knotless anchors were subject to knot failure and locking mechanism failure, respectively.
CLINICAL RELEVANCE
These data support the benefit of SB knotless anchors for anteroinferior labral repair in limiting knot failure typically seen with knotted anchors, perhaps demonstrating that all-suture anchors may have better locking mechanism quality than their PEEK counterparts.