Daniel Wastl, Axel Löwe, Christoph F Dietrich
Journal: Medizinische Klinik, Intensivmedizin und Notfallmedizin 2023;118(4):293-300
PMID: 35691992
BACKGROUND AND AIMS
Targeted ultrasound examinations with a portable ultrasound device ("handheld ultrasound system"; HHUS) have been defined as "echoscopy" by the European Federation of Societies for Ultrasound in Medicine and Biology (EFSUMB). Cardiac scanning with HHUS is feasible. Echoscopy could play a major role in emergency and intensive care medicine, but adequate data on its effectiveness are still lacking. Sonographic examinations in the field of emergency and intensive care medicine can often not be carried out under standardized examination conditions. Thus, the aim of this study is to show that the use of HHUS for echocardiography in emergency medicine is possible and that for this setting HHUS is not inferior to a high-end ultrasound system (HEUS) for detecting cardiac pathologies.
METHODS
The examinations were carried out with a Vscan™ (GE Medical Systems, Solingen, Germany) and a high-end ultrasound device (Acuson X‑300 or X‑700). The examinations were randomized and blinded to two examiners within 30 min. The examinations took place in the intensive care unit, the emergency room and the ambulance service. The results were recorded in an examination sheet.
RESULTS
In all, 93 patients (61 men and 32 women, age 69 ± 14.76 [33-95] years). In 32.6% (30/93) of examinations with HEUS the examination conditions were optimal and in 29.03% (27/93) when the HHUS was used. Of the examinations, 50.08% (31/61) were carried out by both examiners in the same patient position. Using HHUS, the following sensitivity and specificities (respectively) were found: pericardial effusion (73.68%; 96.97%), hemodynamically relevant effusion (50%; 97.67%), right heart strain (90,91%; 96,72%), arrest of the right ventricle (100%; 87,5%), limitation of left ventricular pump function (91.49%; 86.11%), wall movement disorders (WMD, 97.29%; 78.95%), aortic valve sclerosis (42.86%; 86.67%), aortic regurgitation (60%; 95%), mitral valve sclerosis (60%; 100%), mitral reguritation (66.67%; 82.86%), tricuspid valve regurgitation (48%; 81,48%). Measurements of the dimensions of pericardial effusion, the left ventricle, the left atrium and the left ventricular posterior wall each had a positive correlation between the examination with HHUS and HEUS (Κ = 0.45 to 0.91). The diameters determined by HHUS and HEUS for the septum and aortic root, however, correlated negatively (κ = -0.61 to -0.86).
CONCLUSIONS
The use of echoscopy in emergency and intensive care medicine is not inferior to HEUS for detecting defined cardiac pathologies.
© 2022. The Author(s), under exclusive licence to Springer Medizin Verlag GmbH, ein Teil von Springer Nature.
Full Text Sources:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.