Transcatheter Aortic Valve Thrombogenesis: A Foreign Materials Perspective.

Md Tausif Salim, Vahid Sadri, Priya Nair, Daniel Schwaner, Mugdha Apte, Ajit P Yoganathan

Journal: Cardiovascular engineering and technology 2021;12(1):28-36

PMID: 33277684

Abstract

PURPOSE

The initiation of thrombus formation in transcatheter aortic valves (TAVs) is not well understood. The foreign material components of a TAV may play a key role in TAV thrombogenesis. The goal of this study was to evaluate the thrombogenic potential of a TAV (entire valve) and its stent (with skirt).

METHODS

Blood was collected from eight human donors with citrate anticoagulation and later reconstituted with calcium chloride. A low-volume steady flow loop (flow rate = 0.8 L/min) was designed to facilitate three separate conditions (experimental duration = 1 h) per donor blood: (1) control (n = 8), (2) stent-with-skirt (leaflets removed from a 23 mm SAPIEN XT valve; n = 8) and (3) entire valve (an intact 23 mm SAPIEN XT valve; n = 8). Samples were collected at the start and end of each experiment. Serum D-Dimer and thrombin-antithrombin (TAT) concentrations were measured as markers of thrombogenicity.

RESULTS

There was no significant change in serum D-Dimer and TAT concentration with time for the control group. An increasing trend in D-Dimer and TAT concentration was observed with time for the stent-with-skirt group. Interestingly, there was a decreasing trend in serum D-Dimer and TAT concentration with time for the entire valve (leaflet dominating) group. Moreover, changes in D-Dimer and TAT concentration were significantly different between the stent-with-skirt and entire valve (leaflet dominating) groups.

CONCLUSION

Stent-with-skirt was found to impart the most prominent thrombogenic effect, indicating the significance of blood-stent and blood-skirt interactions in TAV thrombosis.

Address: School of Chemical and Biomolecular Engineering, Georgia Institute of Technology, 387 Technology Circle NW, Technology Enterprise Park, Suite 200, Atlanta, GA, 30313, USA.; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.; School of Chemical and Biomolecular Engineering, Georgia Institute of Technology, 387 Technology Circle NW, Technology Enterprise Park, Suite 200, Atlanta, GA, 30313, USA. [email protected].; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA. [email protected].

Link outs

Subscription / membership required

Bant logo

© Copyright 2026, Nutrition Evidence

NED wishes to thank the following organisations for their support: