Quantification of Myocardial Creatine and Triglyceride Content in the Human Heart: Precision and Accuracy of in vivo Proton Magnetic Resonance Spectroscopy.

Adrianus J Bakermans, S Matthijs Boekholdt, Dylan K de Vries, Yolan J Reckman, Emile S Farag, Paul de Heer, Laween Uthman, Simone W Denis, Coert J Zuurbier, Riekelt H Houtkooper, David R Koolbergen, Jolanda Kluin, R Nils Planken, Hildo J Lamb, Andrew G Webb, Gustav J Strijkers, Daniel A Beard, Jeroen A L Jeneson, Aart J Nederveen

Journal: Journal of magnetic resonance imaging : JMRI 2021;54(2):411-420

PMID: 33569824

Abstract

BACKGROUND

Proton magnetic resonance spectroscopy ( H-MRS) of the human heart is deemed to be a quantitative method to investigate myocardial metabolite content, but thorough validations of in vivo measurements against invasive techniques are lacking.

PURPOSE

To determine measurement precision and accuracy for quantifications of myocardial total creatine and triglyceride content with localized H-MRS.

STUDY TYPE

Test-retest repeatability and measurement validation study.

SUBJECTS

Sixteen volunteers and 22 patients scheduled for open-heart aortic valve replacement or septal myectomy.

FIELD STRENGTH/SEQUENCE

Prospectively ECG-triggered respiratory-gated free-breathing single-voxel point-resolved spectroscopy (PRESS) sequence at 3 T.

ASSESSMENT

Myocardial total creatine and triglyceride content were quantified relative to the total water content by fitting the H-MR spectra. Precision was assessed with measurement repeatability. Accuracy was assessed by validating in vivo H-MRS measurements against biochemical assays in myocardial tissue from the same subjects.

STATISTICAL TESTS

Intrasession and intersession repeatability was assessed using Bland-Altman analyses. Agreement between H-MRS measurements and biochemical assay was tested with regression analyses.

RESULTS

The intersession repeatability coefficient for myocardial total creatine content was 41.8% with a mean value of 0.083% ± 0.020% of the total water signal, and 36.7% for myocardial triglyceride content with a mean value of 0.35% ± 0.13% of the total water signal. Ex vivo myocardial total creatine concentrations in tissue samples correlated with the in vivo myocardial total creatine content measured with H-MRS: n = 22, r = 0.44; P < 0.05. Likewise, ex vivo myocardial triglyceride concentrations correlated with the in vivo myocardial triglyceride content: n = 20, r = 0.50; P < 0.05.

DATA CONCLUSION

We validated the use of localized H-MRS of the human heart at 3 T for quantitative assessments of in vivo myocardial tissue metabolite content by estimating the measurement precision and accuracy.

LEVEL OF EVIDENCE

2 TECHNICAL EFFICACY STAGE: 2.

© 2021 The Authors. Journal of Magnetic Resonance Imaging published by Wiley Periodicals LLC. on behalf of International Society for Magnetic Resonance in Medicine.

Address: Department of Radiology and Nuclear Medicine, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Cardiology, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Experimental Cardiology, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Cardiothoracic Surgery, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Radiology and Nuclear Medicine, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; C.J. Gorter Center for High Field MR, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.; Laboratory of Experimental Intensive Care and Anesthesiology, Department of Anesthesiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Laboratory Genetic Metabolic Diseases, Amsterdam Gastroenterology Endocrinology Metabolism, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.; C.J. Gorter Center for High Field MR, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.; Biomedical Engineering and Physics, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.; Department of Molecular and Integrative Physiology, University of Michigan, Ann Arbor, Michigan, USA.; Neuroimaging Center, Department of Neuroscience, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.