Nabil Mehta, Deborah J Hall, Robin Pourzal, Grant E Garrigues
Journal: JBJS reviews 2021;8(9):e1900212
PMID: 32890047
The materials that are used in total shoulder arthroplasty (TSA) implants have been carefully chosen in an attempt to minimize hardware-related complications. The 2 main metal alloys used in TSA implants are Ti-6Al-4V (titanium-aluminum-vanadium) and CoCrMo (cobalt-chromium-molybdenum). Ti alloys are softer than CoCr alloys, making them less wear-resistant and more susceptible to damage, but they have improved osseointegration and osteoconduction properties. Although controversial, metal allergy may be a concern in patients undergoing TSA and may lead to local tissue reaction and aseptic loosening. Numerous modifications to polyethylene, including cross-linking, minimizing oxidation, and vitamin E impregnation, have been developed to minimize wear and reduce complications. Alternative bearing surfaces such as ceramic and pyrolytic carbon, which have strong track records in other fields, represent promising possibilities to enhance the strength and the durability of TSA prostheses.