Pulmonary perfusion by iodine subtraction maps CT angiography in acute pulmonary embolism: comparison with pulmonary perfusion SPECT (PASEP trial).

Brieg Dissaux, Pierre-Yves Le Floch, Philippe Robin, David Bourhis, Francis Couturaud, Pierre-Yves Salaun, Michel Nonent, Pierre-Yves Le Roux

Journal: European radiology 2021;30(9):4857-4864

PMID: 32279113

Abstract

OBJECTIVE

To assess the diagnostic accuracy of iodine map computed tomography pulmonary angiography (CTPA), for segment-based evaluation of lung perfusion in patients with acute pulmonary embolism (PE), using perfusion single-photon emission CT (SPECT) imaging as a reference standard.

METHODS

Thirty participants who have been diagnosed with acute pulmonary embolism on CTPA underwent perfusion SPECT/CT within 24 h. Perfusion SPECT and iodine map were independently interpreted by 2 nuclear medicine physicians and 2 radiologists. For both modalities, each segment was classified as normoperfused or hypoperfused, as defined by a perfusion defect of more than 25% of a segment. The primary end point was the diagnostic accuracy (sensitivity and specificity) of iodine map for segment-based evaluation of lung perfusion, using perfusion SPECT imaging as a reference standard. Following blinded interpretation, a retrospective explanatory analysis was performed to determine potential causes of misinterpretation.

RESULTS

The median time between CTPA with iodine maps and perfusion SPECT was 14 h (range 2-23 h). A total of 597 segments were analyzed. Sensitivity and specificity of iodine maps with CTPA for the detection of segmental perfusion defects were 231/284 = 81.3% (95% CI 76.4 to 85.4%) and 247/313 = 78.9% (95% CI 74.1 to 83.1%), respectively. In retrospect, false results were explained in 48.7%.

CONCLUSION

Iodine map CTPA showed promising results for the assessment of pulmonary perfusion in patients with acute PE, with sensitivity of 81.3% and specificity of 78.9%, respectively. Recognition of typical pitfalls such as atelectasis, fissures, or beam-hardening artifacts may further improve the accuracy of the test.

KEY POINTS

• Sensitivity and specificity of iodine subtraction maps for the detection of segmental perfusion defects were 81.3% (95% CI 76.4 to 85.4%) and 78.9% (95% CI 74.1 to 83.1%), respectively. • Recognition of typical pitfalls such as atelectasis, fissures, or beam-hardening artifacts may further improve the diagnostic accuracy of the test.

Address: Radiology Department, University Hospital, Brest, France.; EA 3878 GETBO IFR 148, Brest, France.; Université de Bretagne Occidentale, Brest, France.; Nuclear Medicine Department, University Hospital, Brest, France.; Pneumology Department, University Hospital, Brest, France.