Validation of Glomerular Filtration Rate Measurement with Blood Sampling from the Injection Site.

Bethelhem W Mekonnen, James M Warwick, Jennifer L Holness

Journal: Journal of nuclear medicine technology 2020;47(4):305-308

PMID: 31182668

Abstract

Measurement of glomerular filtration rate (GFR) from the plasma clearance of a radionuclide-labeled tracer is reliable and accurate. However, to avoid contamination of the blood samples with radioactivity remaining at the injection site, venepuncture at 2 or more sites is required: one for tracer administration and the others for blood sampling. This requirement is uncomfortable for patients, particularly when venous access is difficult. The objective of this study was to validate the use of a single site of venous access in combination with injection site imaging, for GFR measurement. Twenty-two adults (≥18 y) who were referred for GFR determination were included prospectively. GFR was measured from the plasma clearance of Tc-diethylenetriaminepentaacetic acid according to international guidelines. After administration of the tracer through an intravenous cannula, a 60-s static image of the injection site was acquired. A second intravenous cannula was inserted into the contralateral arm. Venous blood samples were collected at 2, 3, and 4 h after administration of the radiotracer from both the injection site (experimental) and the contralateral arm (conventional). GFR was calculated using slope-intercept and single-sample methods. The median conventional and experimental plasma counts (decay- and background-corrected) were compared for the 2-, 3-, and 4-h venous samples. Conventional and experimental GFRs were then compared, with a more than 10% difference between conventional and experimental GFRs being regarded as significant. Four individuals had visible residual activity at the injection site. The median 2-h counts differed significantly between the conventional and experimental sampling sites ( = 0.007), whereas no significant difference was found at 3 or 4 h. When there was a clear injection site image, the difference between the experimental and conventional GFRs was more than 10% in 1 case for single-sample GFR but less than 8% in all cases for slope-intercept GFR. In cases with clear injection site images, slope-intercept GFR calculated after injection site blood sampling showed no clinically significant difference from conventional contralateral-arm sampling.

© 2019 by the Society of Nuclear Medicine and Molecular Imaging.

Address: Nuclear Medicine Division, Department of Medical Imaging and Clinical Oncology, Stellenbosch University, Cape Town, South Africa [email protected].; Nuclear Medicine Division, Department of Medical Imaging and Clinical Oncology, Stellenbosch University, Cape Town, South Africa.
Bant logo

© Copyright 2026, Nutrition Evidence

NED wishes to thank the following organisations for their support:

We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.