Diagnostic performance of myocardial perfusion imaging with conventional and CZT single-photon emission computed tomography in detecting coronary artery disease: A meta-analysis.

Valeria Cantoni, Roberta Green, Wanda Acampa, Emilia Zampella, Roberta Assante, Carmela Nappi, Valeria Gaudieri, Teresa Mannarino, Renato Cuocolo, Eugenio Di Vaia, Mario Petretta, Alberto Cuocolo

Journal: Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 2022;28(2):698-715

PMID: 31089962

Abstract

BACKGROUND

We performed a meta-analysis to compare the diagnostic performance of conventional SPECT (C-SPECT) and cadmium-zinc-telluride (CZT)-SPECT systems in detecting angiographically proven coronary artery disease (CAD).

METHODS

Studies published between January 2000 and February 2018 were identified by database search. We included studies assessing C-SPECT or CZT-SPECT as a diagnostic test to evaluate patients for the presence of CAD, defined as at least 50% diameter stenosis on invasive coronary angiography. A study was eligible regardless of whether patients were referred for suspected or known CAD.

RESULTS

We identified 40 eligible articles (25 C-SPECT and 15 CZT-SPECT studies) including 7334 patients (4997 in C-SPECT and 2337 in CZT-SPECT studies). The pooled sensitivity and specificity were 85% and 66% for C-SPECT and 89% and 69% for CZT-SPECT imaging studies. The area under the curve was slightly higher for CZT-SPECT (0.89) compared to C-SPECT (0.83); accordingly, the summary diagnostic OR was 17 for CZT-SPECT and 11 for C-SPECT. The accuracy of the two tests slightly differs between C-SPECT and CZT-SPECT (chi-square 11.28, P < .05). At meta-regression analysis, no significant association between both sensitivity and specificity and demographical and clinical variables considered was found for C-SPECT and CZT-SPECT studies.

CONCLUSIONS

C-SPECT and CZT-SPECT have good diagnostic performance in detecting angiographic proven CAD, with a slightly higher accuracy for CZT-SPECT. This result supports the use of the novel gamma cameras in clinical routine practices also considering the improvements in acquisition time and radiation exposure reduction.

Address: Department of Advanced Biomedical Sciences, University Federico II, Via Pansini 5, 80131, Naples, Italy.; Institute of Biostructure and Bioimaging, National Council of Research, Naples, Italy.; Department of Translational Medical Sciences, University Federico II, Naples, Italy.; Department of Advanced Biomedical Sciences, University Federico II, Via Pansini 5, 80131, Naples, Italy. [email protected].