Fast kVp-switching dual energy contrast-enhanced thorax and cardiac CT: A phantom study on the accuracy of iodine concentration and effective atomic number measurement.

Antonios E Papadakis, John Damilakis

Journal: Medical physics 2018;44(9):4724-4735

PMID: 28658505

Abstract

PURPOSE

To assess the effect of vessel diameter and exposure parameters on the estimation accuracy of concentration and effective atomic number (Z ) of iodine (I) in contrast-enhanced thorax and cardiac dual-energy CT using a modern fast kVp-switching CT scanner.

METHODS

A standard semi-anthropomorphic cardiac CT phantom devised to simulate the human chest at three different body habitus i.e., medium-sized, large-sized, and obese, was scanned using a fast kVp-switching Revolution-GSI GE CT scanner. Five cylindrical, 10 mm diameter, vials were filled with solutions prepared by diluting I contrast at five concentrations (2.5, 5, 10, 15, and 20 mg I/ml). To simulate small vessels, pipette tips with a diameter ranging from 5 mm to 0.5 mm were employed. The vials and pipette tips were accommodated within the semi-anthropomorphic phantom. CT acquisitions were performed in the fast kVp-switching dual-energy mode at six different CTDI values. Acquisitions were also performed at 80, 100, 120, and 140 kVp. Images were acquired at 64 × 0.625 mm beam collimation and reconstructed at 2.5 mm using all available reconstruction filter kernels. Virtual monochromatic spectral (VMS) images, iodine concentration (I ), and Z maps were reconstructed. Hounsfield unit as a function of energy (HU ) in VMS and single-kVp (HU ), I and Z were measured at each CTDI . The effect of vessel diameter on I and Z was investigated. Measured HU and Z were compared to theoretically estimated values and I were compared to nominal (I ) values.

RESULTS

In 10 mm diameter vessels, HU values were accurate to 18% for the medium-sized, 22% for the large-sized and 39% for the obese phantoms. I was underestimated by up to 10% for the medium-sized, 26% for the large-sized and 33% for the obese phantom. I error decreased with increasing CTDI from ±0.799 mg/ml at 8.61 mGy to ±0.082 mg/ml at 32.01 mGy. The percentage difference between measured and theoretically estimated Z ranged from -3.9% to -14.5%. In pipette tip vessels, I was found to depend on the kernel employed. At the standard kernel, I , for I = 20 mg/ml, was reduced with vessel diameter from 19.25 ± 0.39 mg/ml, at 10 mm, to 2.52 ± 0.31 mg/ml, at 1 mm. Linear regression between I and I resulted in I /I factors of 0.925 for 5 mm, 0.815 for 4 mm, 0.651 for 3 mm, 0.377 for 2 mm, and 0.129 for 1 mm vessel diameter. Measured Z values were underestimated when vessel diameter was decreased from 5 mm to 1 mm by 27% for the 20 mg I/ml and 21% for the 2.5 mg I/ml.

CONCLUSIONS

HU , I , and Z depend on several parameters such as body size, vessel size, exposure parameters, and reconstruction kernel. The limiting spatial resolution of the CT system results in considerable underestimation of HU , I , and Z in vessels smaller than 5 mm diameter. The underestimation of I uptake may be experimentally corrected, if the diameter of the investigated vessel is measured and the correction factors produced in this study are employed.

© 2017 American Association of Physicists in Medicine.