The Assessment of Carbon Dioxide Automated Angiography in Type II Endoleaks Detection: Comparison with Contrast-Enhanced Ultrasound.

Chiara Mascoli, Gianluca Faggioli, Enrico Gallitto, Vincenzo Vento, Giuseppe Indelicato, Rodolfo Pini, Andrea Vacirca, Andrea Stella, Mauro Gargiulo

Journal: Contrast media & molecular imaging 2019;2018():7647165

PMID: 29780288

Abstract

INTRODUCTION

Iodinated contrast media completion angiography (ICM-A) may underestimate the presence of type II endoleak (ELII) after endovascular aortic repair (EVAR), particularly if they are at low flow. Contrast-enhanced ultrasound (CEUS) has been proposed as the gold standard in ELII detection during EVAR follow-up. Intraprocedural carbon dioxide (CO) angiography has been shown to be useful in this setting; however no comparative studies including these three techniques are currently available. Our aim was to investigate the accuracy of a new automated CO angiographic (CO-A) system in the detection of ELII, by comparing it with ICM-A and CEUS.

METHODS

A series of consecutive patients undergoing EVAR for abdominal aortic aneurysm (AAA) were enrolled and submitted to ICM-A and CO-A during the procedure. The iodinated contrast media were delivered through an automatic injector connected to a pigtail catheter in the suprarenal aorta. CO was delivered through a recently available automatic injector connected to a 10 F sheath positioned in the external iliac artery. All patients were blindly evaluated by CEUS within postoperative day 1. The ICM-A and CO-A ability to detect ELII was compared with that of CEUS through Cohen's concordance Index ().

RESULTS

Twenty-one patients were enrolled in the study. One (5%), seven (33%), and four (19%) ELII were detected by ICM-A, CO-A, and CEUS, respectively. The only ELII detected by ICM-A was also detected by CO-A and CEUS. Three cases of ELII detected by CO-A were not detected by CEUS. All ELII detected by CEUS were visualized by CO-A. CEUS and ICM-A showed a poor agreement (Cohen's : 0.35) while CEUS and CO-A showed a substantial agreement (Cohen's : 0.65) for ELII detection.

CONCLUSION

CO-A is safe and effective method for ELII detection in EVAR, with a significantly higher agreement with CEUS if compared with ICM-A. This trial is registered with 155/2015/U/Oss.

Address: Vascular Surgery, DIMES, University of Bologna, Policlinico S. Orsola-Malpighi, Bologna, Italy.