Comparison of self-expanding and balloon-expandable transcatheter aortic valves morphology and association with paravalvular regurgitation: Evaluation using multidetector computed tomography.

João G Almeida, Sara M Ferreira, Paulo Fonseca, Tiago Dias, Cláudio Guerreiro, Ana Barbosa, Pedro Teixeira, Mónica Carvalho, Wilson Ferreira, Nuno D Ferreira, Pedro Braga, José Ribeiro, Vasco G Ribeiro

Journal: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions 2019;92(3):533-541

PMID: 29105953

Abstract

OBJECTIVES

Compare final morphology of self-expanding and balloon-expandable prosthesis and association with paravalvular regurgitation (PVR).

BACKGROUND

PVR after transcatheter aortic valve replacement (TAVR) remains a frequent complication. A better understanding of the prosthesis geometry may be important to improve selection of the best device for each case and possibly reduce the rates of PVR.

METHODS

Retrospective study including patients consecutively submitted to transcatheter aortic valve replacement: August/2007-October/2016. Three months after the procedure a multidetector computed tomography (MDCT) was performed to assess prosthesis geometry: dimensions, eccentricity, and expansion.

RESULTS

A total of 147 individuals were included (mean age of 78.8 ± 6.7 and 50.3% males), 57% treated with a self-expanding prosthesis. On the postprocedure MDCT, the self-expanding group had higher eccentricity index (15.0 vs. 7.1%, p < .001) and lower expansion (68.3 vs. 82.8%, p < .001). In that group, the volume of calcium of landing zone had a significant correlation with eccentricity index and under-expansion. Patients with ≥mild PVR presented higher eccentricity (12.6 vs. 7.9%, p < .001) and lower expansion (68 vs. 75%, p = .012). Eccentricity index and landing zone calcium volume were independent predictors of PVR.

CONCLUSIONS

Self-expanding prosthesis have greater eccentricity and under-expansion. Calcium burden exerts more influence in the final morphology of that type of valve. Calcification and eccentricity are associated with the development of PVR. These factors should be considered in the selection of the most appropriate type of prosthesis for each scenario.

© 2017 Wiley Periodicals, Inc.

Address: Department of Cardiology, Gaia/Espinho Hospital Center, Rua Conceição Fernandes, Vila Nova de Gaia, 4434-502, Portugal.; Department of Cardiology, Divino Espirito Santo Hospital, Ponta Delgada, Portugal.