A I Rappaport, C D Karakochuk, K C Whitfield, K M Kheang, T J Green
Journal: International journal of laboratory hematology 2017;39(1):95-100
PMID: 27997764
INTRODUCTION
Hemoglobin (Hb) concentration is often measured in global health and nutrition surveys to determine anemia prevalence using a portable hemoglobinometer such as the Hemocue® Hb 201+. More recently, a newer model was released (Hemocue Hb 301) utilizing slightly different methods to measure Hb as compared to the older model. The objective was to measure bias and concordance between Hb concentrations using the Hemocue Hb 301 and Hb 201+ models in a rural field setting.
METHODS
Hemoglobin (Hb) concentration was measured using one finger prick of blood (approximately 10 μL) from 175 Cambodian women (18-49 years) using three Hemocue Hb 201+ and three Hb 301 machines. Bias and concordance were measured and plotted.
RESULTS
Overall, mean ± SD Hb concentration was 116 ± 13 g/L using the Hb 201+ and 118 ± 12 g/L using the Hb 301; and anemia prevalence (Hb < 120 g/L) was 58% (n = 102) and 58% (n = 101), respectively. Overall bias ± SD was 2.0 ± 10.5 g/L and concordance (95% CI) was 0.63 (0.54, 0.72). Despite the 2 g/L bias detected between models, anemia prevalence was very similar in both models.
CONCLUSIONS
The two models measured anemia prevalence comparably in this population of women in rural Cambodia.
© 2016 John Wiley & Sons Ltd.