Jonathan Dugernier, Michel Hesse, Rita Vanbever, Virginie Depoortere, Jean Roeseler, Jean-Bernard Michotte, Pierre-François Laterre, François Jamar, Gregory Reychler
Journal: Pharmaceutical research 2017;34(2):290-300
PMID: 27822851
PURPOSE
To compare in vivo the total and regional pulmonary deposition of aerosol particles generated by a new system combining a vibrating-mesh nebulizer with a specific valved holding chamber and constant-output jet nebulizer connected to a corrugated tube.
METHODS
Cross-over study comparing aerosol delivery to the lungs using two nebulizers in 6 healthy male subjects: a vibrating-mesh nebulizer combined with a valved holding chamber (Aerogen Ultra®, Aerogen Ltd., Galway, Ireland) and a jet nebulizer connected to a corrugated tube (Opti-Mist Plus Nebulizer®, ConvaTec, Bridgewater, NJ). Nebulizers were filled with diethylenetriaminepentaacetic acid labelled with technetium-99 m (Tc-DTPA, 2 mCi/4 mL). Pulmonary deposition of Tc-DTPA was measured by single-photon emission computed tomography combined with a low dose CT-scan (SPECT-CT).
RESULTS
Pulmonary aerosol deposition from SPECT-CT analysis was six times increased with the vibrating-mesh nebulizer as compared to the jet nebulizer (34.1 ± 6.0% versus 5.2 ± 1.1%, p < 0.001). However, aerosol penetration expressed as the three-dimensional normalized ratio of the outer and the inner regions of the lungs was similar between both nebulizers.
CONCLUSIONS
This study demonstrated the high superiority of the new system combining a vibrating-mesh nebulizer with a valved holding chamber to deliver nebulized particles into the lungs as comparted to a constant-output jet nebulizer with a corrugated tube.