The Feasibility of a Completely Automated Total IV Anesthesia Drug Delivery System for Cardiac Surgery.

Cedrick Zaouter, Thomas M Hemmerling, Romain Lanchon, Emanuela Valoti, Alain Remy, Sébastien Leuillet, Alexandre Ouattara

Journal: Anesthesia and analgesia 2017;123(4):885-93

PMID: 27644009

Abstract

BACKGROUND

In this pilot study, we tested a novel automatic anesthesia system for closed-loop administration of IV anesthesia drugs for cardiac surgical procedures with cardiopulmonary bypass. This anesthesia drug delivery robot integrates all 3 components of general anesthesia: hypnosis, analgesia, and muscle relaxation.

METHODS

Twenty patients scheduled for elective cardiac surgery with cardiopulmonary bypass were enrolled. Propofol, remifentanil, and rocuronium were administered using closed-loop feedback control. The main objective was the feasibility of closed-loop anesthesia defined as successful automated cardiac anesthesia without manual override by the attending anesthesiologist. Secondary qualitative observations were clinical and controller performances. The clinical performance of hypnosis control was the efficacy to maintain a bispectral index (BIS) of 45. To evaluate the hypnosis performance, BIS values were stratified into 4 categories: "excellent," "good," "poor," and "inadequate" hypnosis control defined as BIS values within 10%, ranging from 11% to 20%, ranging from 21% to 30%, or >30% of the target value, respectively. The clinical performance of analgesia was the efficacy to maintain NociMap values close to 0. The analgesia performance was assessed classifying the NociMap values in 3 pain control groups: -33 to +33 representing excellent pain control, -34 to -66 and +34 to +66 representing good pain control, and -67 to -100 and +67 to +100 representing insufficient pain control. The controller performance was calculated using the Varvel parameters.

RESULTS

Robotic anesthesia was successful in 16 patients, which is equivalent to 80% (97.5% confidence interval [CI], 53%-95%) of the patients undergoing cardiac surgery. Four patients were excluded from the final analysis because of technical problems with the automated anesthesia delivery system. The secondary qualitative observations revealed that the clinical performance of hypnosis allowed an excellent and good control during 70% (97.5% CI, 63%-76%) of maintenance time and an insufficient clinical performance of analgesia for only 3% (97.5% CI, 1%-6%) of maintenance time.

CONCLUSIONS

The completely automated closed-loop system tested in this investigation could be used successfully and safely for cardiac surgery necessitating cardiopulmonary bypass. The results of the present trial showed satisfactory clinical performance of anesthesia control.

Address: From the *CHU de Bordeaux, Service d Anesthésie-Réanimation II, Bordeaux, France; †Division of Experimental Surgery, Department of Anaesthesia, Arnold and Blema Steinberg Simulation Centre, McGill University, Montreal, Canada; ‡Servizio di Anestesia Rianimazione e terapia intensiva, Università degli Studi di Pavia, Dip. Scienze Clinico-Chirurgiche Diagnostiche e Pediatriche, Unità di Anestesiologia - Fondazione IRCCS Policlinico San Matteo, Pavia, Italy; and §Biofortis, Mérieux NutriSciences Company, Saint-Herblain, France.