Arnaud Potié, Caroline Prégardien, Thierry Pirotte, Xavier Stephenne, Isabelle Scheers, Catherine Wanty, Françoise Smets, Etienne Sokal, Francis Veyckemans
Journal: Paediatric anaesthesia 2017;26(6):649-54
PMID: 27111823
AIMS
The aim of this study was to evaluate the usability and safety of the Explorer Endoscopy Mask(®) (EM) as an alternative to endotracheal intubation in children undergoing elective esogastroduodenoscopy (EGD) under general anesthesia (GA).
METHODS
This study was a retrospective observational study. The study was undertaken at the pediatric digestive endoscopy suite in the Cliniques universitaires Saint-Luc, Brussels, Belgium. We retrospectively analyzed the occurrence of minor and major airway-related adverse effects during pediatric EGD procedures performed under GA with the EM between June 2014 and March 2015.
RESULTS
During the study period, 173 patients underwent EGD. Their mean age was 8.4 years (median: 9.1 years, range 4 months to 16 years). Mean duration of endoscopy (from insertion to removal of the endoscope) was 12.6 min (median: 12 min, range 3-47 min). The use of EM was uneventful in 159 (92%) cases. There were 24 airway-related adverse events in 14 children. Hypoxemia (SpO2 <90%) (13 events, 7.5%) was the most commonly encountered complication followed by laryngo- or bronchospasm (five events, 2.89%), cough (five events, 2.89%), and intubation (one event, 0.58%). No cases of regurgitation/aspiration were observed.
CONCLUSIONS
Our data support the EM use in pediatric EGD. There were few transient respiratory adverse events which were easily solved with minor interventions.
© 2016 John Wiley & Sons Ltd.