Philippe Meyer, Anne-Marie Makhlouf, Louis P Mondouagne Engkolo, Florence Trentaz, Ronan Thibault, Claude Pichard, Haran Burri
Journal: JPEN. Journal of parenteral and enteral nutrition 2018;41(6):981-985
PMID: 26903305
BACKGROUND
Current guidelines warn of potential electromagnetic interferences (EMI) when using bioelectrical impedance analysis (BIA) to measure body composition in patients equipped with implantable cardioverter-defibrillators (ICDs). We aimed to test the occurrence of EMI in a setting where this risk was experimentally maximized.
MATERIALS AND METHODS
Outpatients scheduled for routine ICD controls simultaneously underwent a BIA measurement using an electrical current of 0.8 mAmp at frequencies from 5-100 kHz. ICD sensitivity levels were set to maximum levels while therapies were temporarily inactivated. The device electrograms were monitored in real time to detect sensed and/or visible EMI during BIA measurement.
RESULTS
A total of 63 patients equipped with single-chamber (n = 13), dual-chamber (n = 18), or biventricular (n = 32) ICDs from 5 major manufacturers were included. No EMI were detected by the ICDs in these patients, nor were any artifacts visualized during real-time electrogram recordings.
CONCLUSION
BIA can be safely performed in patients equipped with ICDs without cardiac monitoring. Current guidelines should be updated accordingly.