AIMS
Sensory neuropathy is central to the development of painful neuropathy, and foot ulceration in patients with diabetes. Currently, available QST devices take considerable time to perform and are expensive. NerveCheck is the first inexpensive ($500), portable QST device to perform both vibration and thermal testing and hence evaluate diabetic peripheral neuropathy (DPN). This study was undertaken to establish the reproducibility and diagnostic validity of NerveCheck for detecting neuropathy.
METHODS
130 subjects (28 with DPN, 46 without DPN and 56 control subjects) underwent QST assessment with NerveCheck; vibration perception and thermal testing. DPN was defined according to the Toronto criteria.
RESULTS
NerveCheck's intra correlation coefficient for vibration, cold and warm sensation testing was 0.79 (95% LOA: -4.20 to 6.60), 0.86 (95% LOA: -1.38 to 2.72) and 0.71 (95% LOA: -2.36 to 3.83), respectively. The diagnostic accuracy (AUC) for vibration, cold and warm sensation testing was 86% (SE: 0.038, 95% CI 0.79-0.94), 79% (SE: 0.058, 95% CI 0.68-0.91) and 72% (SE: 0.058, 95% CI 0.60-0.83), respectively.
CONCLUSIONS
This study shows that NerveCheck has good reproducibility and comparable diagnostic accuracy to established QST equipment for the diagnosis of DPN.
Copyright © 2016 The Authors. Published by Elsevier Ireland Ltd.. All rights reserved.