Jing Zou, Juha Koivisto, Jaakko Lähelmä, Antti Aarnisalo, Jan Wolff, Ilmari Pyykkö
Journal: The Annals of otology, rhinology, and laryngology 2015;124(6):466-73
PMID: 25533506
OBJECTIVE
To evaluate the impact of tube voltage, tube current, pulse number, and magnification factor on the image quality of a novel experimental set-up and the corresponding radiation.
MATERIALS AND METHODS
Six human temporal bones with cochlear implant were imaged using various tube voltages, tube currents, pulse numbers, and magnification. The effect of radiation was evaluated using a metaloxide semiconductor field-effect transistor (MOSFET) dosimeter device on an anthropomorphic RANDO RAN102 male head phantom. A copper and aluminum combination filter was used for hardware filtration.
RESULTS
Overall, 900 frames, 11 mA, and 88 kV provided the best image quality. In temporal bones imaged with the optimized parameters, the cochlea, osseous spiral lamina, modiolus, stapes, round window niche, and oval window landmarks were demonstrated with anatomic structures still fully assessable in all parts and acceptable image quality. The most dominant contributor to the effective dose was bone marrow (36%-37 %) followed by brain (34%-36%), remainder tissues (12%), extra-thoracic airways (7%), and oral mucosa (5%).
CONCLUSIONS
By increasing the number of frames, the image quality of the inner ear details obtained using the novel cone-beam computed tomography improved.
© The Author(s) 2014.