Joerg Neuzner, Thomas Dietze, Robert Paliege, Michael Möller, Giovanni Saeed, Rainer Gradaus
Journal: Cardiology journal 2016;22(2):201-5
PMID: 25299502
BACKGROUND
In left atrial appendage (LAA) closure, the correct sizing of the implantable devices is crucial. Data on the time-dependent changes in the shape and positioning of LAA occlusion devices are missing. We analyzed the results of 33 consecutive patients after implantation of an Amplatzer™ Cardiac Plug (ACP) LAA closure device to get more information on the optimal device sizing during implantation.
METHODS AND RESULTS
Thirty-three consecutive patients were enrolled in this observational study. ACP implantation was guided by fluoroscopy and three dimensional transesophageal echocardiography (3-D TEE). Device sizing was based on the largest measured diameter of the intended landing zone adding 2-4 mm of device oversizing. Fluoroscopies were performed at 1 day after, and after 3 months, control 3-D TEE was performed 3 months after implantation. The stability of device positioning and shape was matched with the results of 3-D TEE. Patients' mean age was 70.2 ± 8 years; mean CHA2DS2VASc score was 3.8 ± 1.1. According to the manufacture's classification, the post-implant degree of compression of the device-lobe was classified in three categories 1) undercompression "square-like shape" (1 patient); 2) optimal compression "tire-like shape" (20 patients), 3) overcompression "strawberry-like shape" (12 patients). Changes in the degree of device compression by more than one classification class occurred in 18/33 of our patients. A complete loss of device compression ("square-like shape") was observed in 9 patients. Despite the changes in device compression, a complete closure of the LAA was achieved in 32/33 patients.
CONCLUSIONS
There is a temporal change in shape and positioning of the ACP within 3 months after implantation. A late decompression of the ACP lobe was observed in 61% of our patients, leading to a complete loss in device compression in 27%. This observation may be the rationale for a higher degree of ACP oversizing during implantation.
Full Text Sources:
© Copyright 2026, Nutrition Evidence
We use cookies to improve your experience and analyze site traffic with Google Analytics. By continuing to use our site, you agree to our use of cookies. Learn more.