High speed insertion of bone fracture fixation pins: a finite element penetration model with experimental comparisons.

Matthew P Prygoski, Timotius Pasang, Steven R Schmid, Antony J Lozier

Journal: Journal of materials science. Materials in medicine 2015;22(12):2823-32

PMID: 22042459

Abstract

A new method of bone fracture fixation is considered in which small pins/darts are dynamically inserted into bone to prevent translation and rotation at the fracture site. An ABAQUS model was developed to analyze dart penetration in cortical and cancellous bone for varying dart diameter, material, and velocity, and cortical thickness. The method is advocated for bioresorbable darts, so polylactide (PLA) and magnesium are the materials examined in this study. Numerical results showed that magnesium darts can achieve full penetration in bone while suffering little damage. The PLA darts penetrated thin bone well, but substantial deformation was seen as the cortical thickness increased, especially for small diameter darts. As partial validation, prototype PLA fixation darts were fired into cadaveric bone with a custom nailer. As in the model, the PLA darts could penetrate thin cortices but saw gross deformation when impacted against thicker bone.

Address: Department of Aerospace and Mechanical Engineering, University of Notre Dame, Notre Dame, IN 46637, USA.

Link outs

Free resources

Subscription / membership required

Bant logo

© Copyright 2026, Nutrition Evidence

NED wishes to thank the following organisations for their support: