Antonio Moroni, Emanuele Nocco, Martha Hoque, Elena Diremigio, Dario Buffoli, Francesco Cantù, Simona Catalani, Pietro Apostoli
Journal: Archives of orthopaedic and trauma surgery 2012;132(1):123-9
PMID: 21845437
INTRODUCTION
Metal-on-metal total hip arthroplasty (MOM THA) has the advantage of replicating the femoral head size, but the postoperative elevation of serum metal ion levels is a cause for concern. Metal-on-polycarbonate-urethane is a new cushion bearing featuring a large diameter metal head coupled with a polycarbonate-urethane liner.
AIM
The aim of this study was to assess and compare serum cobalt (Co) and chromium (Cr) levels in a group of 15 patients treated with a cushion bearing THA system (Group A) and a group of 15 patients treated with a MOM THA system (Group B) at short-term. At a mean follow-up of 27.3 months (18-35 months), in Group A the median Cr and Co serum levels were significantly lower than in Group B, measuring 0.24 μg/L (0.1-2.1 μg/L) and 0.6 μg/L (0.29-2.3 μg/L) compared to 1.3 μg/L (0.1-9 μg/L, p < 0.001) and 2.9 μg/L (0.85-13.8 μg/L, p < 0.001) respectively.
RESULTS
All patients demonstrated an excellent clinical result, as shown by the Harris and Oxford hip scores. The cushion bearing THA studied in this paper showed clinical outcomes similar to the MOM THA bearing, with the advantage of no significant metal ion elevation in the serum.
CONCLUSION
These findings warrant the continued clinical study of compliant bearing options.