Louis A Cannon, Dean J Kereiakes, Tift Mann, Jeffrey J Popma, Michael R Mooney, Gregory J Mishkel, Tommy C Lee, B Hadley Wilson, Thomas D Stuckey, Steven Orlow, Thomas McGarry, Michael E Ring, Mirle A Kellett, Paul Underwood, Keith D Dawkins
Journal: EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology 2011;6(8):920-7, 1-2
PMID: 21330238
AIMS
Small reference vessel diameter predicts adverse outcomes following coronary stenting. TAXUS Express and TAXUS Liberté paclitaxel-eluting stents (PES) reduce restenosis compared to bare metal stents (BMS) in small diameter vessels. TAXUS Element is a novel thin-strut, platinum chromium stent designed to enhance visibility, conformability, and drug delivery in small diameter vessels.
METHODS AND RESULTS
The PERSEUS Small Vessel (SV) prospective, single-arm, superiority trial evaluates the TAXUS Element PES in 224 subjects with target lesion length≤20 mm and vessel diameter≥2.25 to <2.75 mm, compared to 125 lesion-matched historical Express BMS control subjects from the TAXUS V trial. The primary endpoint was nine-month in-stent late loss. The secondary endpoint was 12-month target lesion failure (TLF) compared to a pre-specified performance goal (PG). Outcomes were analysed with and without propensity-score adjustment. TAXUS Element was superior to the Express BMS for late loss (0.38±0.51 versus 0.80±0.53 mm respectively; P<0.001), and TLF (7.3%) was significantly less than the 19.5% PG (P<0.001). No differences in mortality, myocardial infarction, or stent thrombosis were observed through 12 months. Results were similar after adjustment.
CONCLUSIONS
PERSEUS SV supports the efficacy and safety of the platinum chromium, thin-strut TAXUS Element stent in small coronary vessels.
© Copyright 2026, Nutrition Evidence