Serita Barnard, Edward Leen, Timothy Cooke, Wilson Angerson
Journal: South African medical journal = Suid-Afrikaanse tydskrif vir geneeskunde 2008;98(5):386-91
PMID: 18637311
OBJECTIVE
To determine the diagnostic value of haemodynamic contrast-enhanced ultrasound assessment in benign and malignant breast tissue, using histological examination as the reference standard.
METHODS
An HDI 5000 Phillips ultrasound scanner with microvascular imaging software and 2.5 ml SonoVue (Bracco spa, Milan) contrast was used to scan 50 consecutive patients (32 malignant and 18 benign, 49 with histologically confirmed breast lesions). Time-intensity curves of the regions of interest (ROI) placed over the lesional and normal breast tissues were acquired using QLAB software. The area under the curve (AUC), time to peak (TTP), in-flow gradient (IFG) and peak enhancement (PE) were determined in a standardised manner for each focal breast lesion and the control/normal breast. IFG and AUC in the periphery, and PE and AUC in the centre of the lesions, differed significantly between malignant and benign tumours. In a logistical regression model, AUC in the periphery and the ratio of PE at the tumour centre to that of normal breast were independently predictive of malignancy (p < 0.001), achieving a diagnostic sensitivity of 97% and a specificity of 80% in the 41 tumours in which central vessels could be distinguished.
CONCLUSION
Haemodynamic contrast-enhanced ultrasound assessment can be used to distinguish between benign and malignant breast lesions.